MIB Viewer

moduleEventStatus

OBJECT-TYPE
1.3.6.1.4.1.19947.1.3.6.1.9
Access
read-only
Status
current
Sub-identifiers
0
A bit string which shows the current module status. If a bit is set (1), the explanation is satisfied: moduleEventPowerFail (0), Event power fail generated by the MPOD controller in order to ramp down all HV's (ramp speed=1000V/s) moduleEventLiveInsertion (2), Event live insertion to prepare a hot plug of a module moduleEventService (4), Module event: Hardware failure detected (consult iseg Spezialelektronik GmbH). moduleHardwareLimitVoltageNotGood (5), Module Event: Hardware voltage limit is not in proper range, using for HV distributor modules with current mirror only. moduleEventInputError (6), Module event: There was an input error in connection with a module access moduleEventSafetyLoopNotGood (10), Module event: Safety loop is open. moduleEventSupplyNotGood (13), Module event: At least one of the supplies is not good. moduleEventTemperatureNotGood (14), Module event: Temperature was above the permitted threshold (for EDS/EHS temperature above 55 degr.C)
BITS { moduleEventPowerFail (0), moduleEventLiveInsertion (2), moduleEventService (4), moduleHardwareLimitVoltageNotGood (5), moduleEventInputError (6), moduleEventSafetyLoopNotGood (10), moduleEventSupplyNotGood (13), moduleEventTemperatureNotGood (14) }